To ensure that your hardness measurements are accurate, you must calibrate your Rockwell hardness tester. To ensure precise readings, the performance of the machine must be compared to predetermined benchmarks and adjusted accordingly. Here's a general overview of the process:

Preparation:

  1. Gather materials: You will need your tester's handbook, a cleaning brush, certified reference blocks with the appropriate scales and hardness levels, and a Rockwell hardness tester.
  2. Clean the tester: To remove any dirt or debris that can affect accuracy, wipe down the anvil, indenter, and other components with a clean, dry cloth.
  3. Choose the test scales: Choose which Rockwell scales you will be calibrated based on the materials you typically test.

Calibration procedure:

  1. Set the indenter: Check to make sure the right indenter—a steel ball or diamond—is mounted and positioned appropriately on the chosen scale.
  2. Zero the gauge: To zero the dial gauge, use the zeroing screw or knob, depending on your model.
  3. Test the reference blocks: Firmly fasten the anvil to the first reference block of known hardness. Apply the main and preliminary loads for the chosen scale in compliance with the manufacturer's recommendations.
  4. Read and record the hardness value: After the dwell time, release the load and verify the hardness value on the dial gauge. This value ought to be contrasted with the certified hardness of the reference block.
  5. Repeat for other reference blocks: Steps 3–4 should be repeated for each extra reference block until the entire range of hardness that you typically use is covered.
  6. Adjust the indenter (if needed): If the readings consistently deviate from the certified values of the reference blocks, you may need to slightly adjust the indenter position. Consult the manufacturer's instructions for comprehensive adjustment instructions.
  7. Re-test and document: After making any changes, retest the reference blocks and note the updated readings.

Additional tips:

  • Ensure that the reference blocks you utilize are made of the same material or one that is similar to the specimens you test regularly.
  • Perform the calibration within a climate-controlled area that maintains a constant temperature and humidity level.
  • If you frequently use your tester or use it for critical applications, consider having a qualified technician calibrate it professionally regularly.